Configuration (administrator)
•
Read from Backup Database: With this button you can read the values backwards
from the backup database. This is useful if you want to reuse a former configuration.
•
Write Values: Pressing this button will write the current field values to the unit.
Warning: This will overwrite the current values saved in the control unit mem-
ory.
Write to Backup Database: When you press this button you can save the current field
values in the database of the BMPC/2 control program.
i
Changes to the limit values will first take effect when you press the Write Values button.
Once you are certain that the values are suitable you can add them later to the backup
database by pressing the Write to Backup Database button.
12.3.2.2
Heater & APD Settings
In this settings window you can adjust the settings for the magnet heater.
Figure 12.15Tools - Options - Heater & APD Setting
These values are normally entered during the installation and in most cases do not need
to be modified.
•
Heater Mode: It is possible to operate in automatic or manual heater mode. As
default Automatic mode is selected. In automatic mode, internal regulation ensures
that the P2 pressure remains constant in the magnet dewar, based on the values
entered. In manual mode (Automatic must be deselected) the unit will be heated
using a constant heater power.
Summary of Contents for BMPC/2
Page 8: ...8 Z31823_00_05 Table of Contents...
Page 14: ...14 Z31823_00_05 Safety...
Page 20: ...20 Z31823_00_05 Principles of Operation...
Page 42: ...42 Z31823_00_05 Alarm System...
Page 62: ...62 Z31823_00_05 Hardware Overview...
Page 64: ...64 Z31823_00_05 Maintenance and Cleaning...
Page 86: ...86 Z31823_00_05 Troubleshooting Figure 11 10Replacing the Control Unit Fuse Fuse Insert...
Page 122: ...122 Z31823_00_05 Technical Data...
Page 124: ...124 Z31823_00_05 Contact...
Page 126: ...126 Z31823_00_05 Warning Signs...
Page 130: ...130 Z31823_00_05 Figures...
Page 132: ...132 Z31823_00_05 Tables...
Page 135: ......