Planning details
Figure xy: (1) relation for the required input pressure difference (dP) and the corresponding pri-
mary water temperature (T) for the total required cooling capacity of approx.11 kW.
Specifications for the water quality
Specifications for the water quality
Table 6.21: Specifications for the water quality
6.10.1
Connections and installation
The basic system's chiller cabinet is connected to the local cold water system via two ¾“ pres-
sure hoses with hose clamps. If the MRI CryoProbe™ is used, two additional parallel connec-
tions must be planned and provided. The maximum distance between the connection points of
the cold water system of the MR system and the installation site of the chiller cabinet may not
exceed 6 m.
Other equipment options do not require any addition connections to the cold water system.
Attention should be paid to ensure that the required values are checked before installation be-
gins.
Water quality
Specification
pH range
6.5 – 8.2
hardness
5.6 – 11.2 °d
maximum proportion of glycol in the water
30 %
chloride
< 100 mg/l
solid floating particles
< 10 mg/l
maximum particle size
< 400 µm
Summary of Contents for BioSpec 117 AVANCE III
Page 1: ...Innovation with Integrity MRI T Site planning BioSpec 117 16 AVANCE III Version 1 7...
Page 6: ...Contents 6 ID 182463115...
Page 33: ...ID 182463115 33 Planning details Stray field Figure 6 3 Stray field...
Page 72: ...72 ID 182463115 Planning details...
Page 82: ...82 ID 182463115 Delivery and transport...
Page 84: ...84 ID 182463115 Moving or dismantling the MR system...
Page 92: ...Figures 92 ID 182463115...
Page 94: ...Tables 94 ID 182463115...
Page 101: ...ID 182463115 107 Lastpage...
Page 102: ...Bruker Corporation info bruker com www bruker com...