Planning details
Further requirements for the air-conditioning system in the technical room
Further requirements for the space cooling in the technical room
Table 6.19: Further requirements for the space cooling in the technical room
6.9.3
Anesthetic gas extraction
If anesthetic gases are used, attention must be paid to the local regulations on maximum work
place concentrations of anesthetic gases (occupational standards) and these regulations must
be adhered to.
Installation of a separate gas extraction system for anesthetic gases is principally recommend-
ed. The MR system provides a connection point to which it can be connected. Gases are ex-
tracted in the vicinity of the test object and discharged to the gas extraction system via a tube 3
m long with an inner diameter of 25 mm. The connection point to the gas extraction system of
the building is located above the CCM at the service end of the magnet. Using system-supplied
suction pipes a suction capacity of about 15 m
3
/ h of extraction will be required.
Anesthetic gas extraction
Figure 6.21: Anesthetic gas extraction
The following additional cooling performances in the technical room must be taken into account
when selecting options for a standard configuration.
Equipment options: Further requirements for the space
cooling for the technical room
Additional cooling perfor-
mance
Option: High Power gradient amplifier
7 kW
Option: MRI CryoProbe™
1.2 kW
Option: Paralell Transmit Cabinet
2.0 kW
1
Extraction tubes provided on the side of the MR system
2
System for extracting anesthetic gases provided by the owner
Summary of Contents for BioSpec 117 AVANCE III
Page 1: ...Innovation with Integrity MRI T Site planning BioSpec 117 16 AVANCE III Version 1 7...
Page 6: ...Contents 6 ID 182463115...
Page 33: ...ID 182463115 33 Planning details Stray field Figure 6 3 Stray field...
Page 72: ...72 ID 182463115 Planning details...
Page 82: ...82 ID 182463115 Delivery and transport...
Page 84: ...84 ID 182463115 Moving or dismantling the MR system...
Page 92: ...Figures 92 ID 182463115...
Page 94: ...Tables 94 ID 182463115...
Page 101: ...ID 182463115 107 Lastpage...
Page 102: ...Bruker Corporation info bruker com www bruker com...