Securing the Bottom of the Column Assembly
Installation Manual Version 001
BRUKER
21 (167)
Place the pneumatic arm assembly over the top of the column, sliding the two fix-
ture rings (
) over the round column piece. Tighten the rings just
enough to keep the pneumatic arm assembly from sliding down.
Rotate the pneumatic arm assembly (
) until the pincher is directly over
the shim system
). The distance between the bottom of the pincher
(when the cylinder is in the down position) and the top of the shim system should
be 3 cm. Adjust this distance by loosening the fixture rings and raising or lowering
the pneumatic arm assembly. When the correct distance is achieved, tighten the
fixture rings. Raise the pneumatic arm resting ring until it rests firmly against the
bottom fixture ring and tighten it securely.
Figure 4.9. Position of the Pneumatic Arm on the Column
Securing the Bottom of the Column Assembly
4.9
Using a level measuring device, check to see if the round column piece is vertical-
ly level. Move the bottom of the column assembly as needed to adjust the level.
PINCHER
VERTICAL
CYLINDER
FIXTURE
RINGS
PISTON
PNEUMATIC ARM
RESTING RING
MAGAZINE BELT
TOP FASTENING
RING
CABINET
MAGNET
558
90
27
7
440
APPRO
X
39
5
APPROX
PNEUMATIC ARM
Summary of Contents for B-ACS 120
Page 1: ...BRUKER Version Bruker Automated Control Systems B ACS 60 120 Installation Manual 001...
Page 7: ...Installation Manual Version 001 BRUKER 7 167 1 Declaration of Conformity 1...
Page 10: ...10 167 BRUKER Installation Manual Version 001 Introduction...
Page 40: ...40 167 BRUKER Installation Manual Version 001 Operating Instructions...
Page 56: ...56 167 BRUKER Installation Manual Version 001 Special Tools...
Page 60: ...60 167 BRUKER Installation Manual Version 001 Technical Data...
Page 150: ...150 167 BRUKER Installation Manual Version 001...
Page 160: ...160 167 BRUKER Installation Manual Version 001 Figures...
Page 167: ...Installation Manual Version 001 BRUKER 167 167 Lastpage...