Magnetic Stray Fields
H157655_1_008
41
6.2
Vertical Stray Fields
Magnet Type
5.0 mT
(50 G)
3.0 mT
(30 G)
1.0 mT
(10 G)
0.5 mT
(5 G)
0.2 mT
(2 G)
0.1 mT
(1 G)
0.05 mT
(0.5 G)
400/54 Ascend Aeon
0.64
0.70
0.87
1.00
1.24
1.47
1.77
500/54 Ascend Aeon
0.80
0.87
1.06
1.20
1.48
1.75
2.09
600/54 Ascend Aeon
0.90
0.99
1.21
1.40
1.71
2.03
2.45
700/54 Ascend Aeon
0.99
1.09
1.37
1.60
2.01
2.42
2.95
Distances are measured in
axial
direction from magnetic center.
Table 6.4:
Vertical Stray Fields: Standard Bore 54 mm
Magnet Type
5.0 mT
(50 G)
3.0 mT
(30 G)
1.0 mT
(10 G)
0.5 mT
(5 G)
0.2 mT
(2 G)
0.1 mT
(1 G)
0.05 mT
(0.5 G)
400/89 Ascend Aeon
0.79
0.86
1.05
1.20
1.45
1.70
2.02
500/89 Ascend Aeon
0.88
0.97
1.21
1.40
1.73
2.07
2.49
600/89 Ascend Aeon
1.02
1.12
1.38
1.60
1.97
2.34
2.81
700/89 Ascend Aeon
1.56
1.72
2.15
2.50
3.09
3.67
4.41
Distances are measured in
axial
direction from magnetic center.
Table 6.5:
Vertical Stray Fields: Wide Bore 89 mm
Summary of Contents for AVANCE
Page 6: ...Contents vi H157655_1_008 ...
Page 24: ...System Components 24 H157655_1_008 ...
Page 32: ...Magnet Access and Rigging 32 H157655_1_008 ...
Page 38: ...Ceiling Height Requirements 38 H157655_1_008 ...
Page 72: ...Utility Requirements 72 H157655_1_008 ...
Page 82: ...Floor Plan 82 H157655_1_008 ...
Page 96: ...CryoProbe and Other Accessories 96 H157655_1_008 ...
Page 102: ...Contact 102 H157655_1_008 ...
Page 104: ...List of Figures 104 H157655_1_008 ...
Page 109: ...H157655_1_008 109 ...
Page 110: ...Bruker Corporation info bruker com www bruker com Order No H157655 ...