Adjustment of a Chamber on the D8
42
DOC-M88-EXX268_V1_12.2016
3. repeat the scan in a narrower range with a smaller step size
Figure 4.5: More accurate Z-scan to determine the right sample height
u
If you do not have a motorized Z-stage,
1. Open the ratemeter (
COMMANDER | Commander drop-down menu | Ratemeter
)
and open the shutter.
2. Move the sample fully out of the beam.
3. Mark the count rate or save it by clicking the
Save as reference
button.
4. Close the shutter, move the chamber using the manual adjustment screw and check
the count rate again.
5. Repeat until the ratemeter shows 50% of the direct beam intensity.