Appendix / Service Details
126
DOC-M88-EXX268_V1_12.2016
10.2.3
DCS 350 and DHS 1100
CAUTION
Risk of severe burns when touching the PEEK Dome
u
Do not touch the PEEK Dome.
u
Do not touch the PEEK Dome because it breaks very easy under vacuum.
For configuring the DCS 350 and DHS 1100 stages, a slightly different procedure from the
one described in paragraph
HTK-series, TTK 600, XRK 900, and HPC 900 Chambers
needs to be followed. Figures 10.1 until 10.5 are still the same. From there on,
take the following steps.
u
As the dome stages are not mounted in the goniometer center, but on sample stages,
they have to be activated as attachment on the mounted stage.
1. To do this, in
Goniometer | Center | “your mounted stage”
, in the line
AttachmentContainer
, column
Connected to Component
, select
/Device/
Goniometer/Center/SampleAttachments
:
Figure 10.38: Setting the plug for SampleAttachments
For
SerialConnector
you have 6 possibilities to connect the RS232 communication line.