32
ZTKS0175 / Z31991 / 03
Transportation
3.5
Transportation with a Crane
A crane is recommended for lifting the cryostat out of the box.
Approved Persons
: Approved crane operator
Precondition
: The crane must be approved for the transportation weight (see
Appendix, Technical Data
).
Attachment Points
Figure 3.4:
Attachment Points for Lifting Equipment
1.
Exclusively use the marked eyelets as
attachment points for the lifting equipment.
2.
Use
all
eyelets for the lifting equipment.
Figure 3.5:
Instruction Label for Lifting Equipment
3.
Follow the instructions on the label on top of
the cryostat. This label gives important
information about correct attachment and
transportation.
4.
Check for correct fastening of the lifting
equipment before lifting the cryostat.
5.
Make sure that any movement of the crane
is as slowly as possible to avoid any
damage due to acceleration.
6.
Check for correct leveling of the cryostat
while hanging on the crane.
Summary of Contents for 600'54 Ascend ULH
Page 1: ...Version Innovation with Integrity 600 54 Ascend ULH 03 User Manual NMR ...
Page 4: ...4 ZTKS0175 Z31991 03 ...
Page 7: ...7 ZTKS0175 Z31991 03 ...
Page 8: ...8 ZTKS0175 Z31991 03 π ...
Page 10: ...10 ZTKS0175 Z31991 03 Contact ...
Page 34: ...34 ZTKS0175 Z31991 03 Transportation ...
Page 35: ...35 ZTKS0175 Z31991 03 4 Assembling 4 1 Safety Approved Persons Bruker Service only ...
Page 36: ...36 ZTKS0175 Z31991 03 Assembling ...
Page 57: ...57 ZTKS0175 Z31991 03 8 Disassembling 8 1 Safety Approved Persons Bruker Service only ...
Page 58: ...58 ZTKS0175 Z31991 03 Disassembling ...
Page 60: ...60 ZTKS0175 Z31991 03 Appendix ...
Page 62: ...62 ZTKS0175 Z31991 03 Appendix ...
Page 64: ...64 ZTKS0175 Z31991 03 Appendix ...
Page 66: ...66 ZTKS0175 Z31991 03 Appendix ...
Page 68: ...68 ZTKS0175 Z31991 03 Appendix ...
Page 102: ...Bruker Corporation info bruker com www bruker com Order No ZTKS0175 Z31991 03 ...