67
ZTKS0168 / Z31960 / 05
Technical Data MS 300‘89 Ascend
9.7
Temperature Sensors
The temperature sensors (PT 100 and IBT) are used to monitor the temperature of the
magnet during cooling down and warming up the magnet system.
PT 100 Sensor
Measure the resistance with a maximum current of 1 mA.
Temperature
Unit
Resistance
Unit
Room Temperature
293
K
107.8
Ω
273
K
100.0
Ω
250
K
91.0
Ω
200
K
71.1
Ω
150
K
50.9
Ω
100
K
30.0
Ω
Liquid Nitrogen
77
K
20.1
Ω
Table 9.9: Characteristic values of PT 100 sensor
Figure 9.6:
Characteristic curve of PT 100 sensor