Troubleshooting
6.3
Troubleshooting Work
6.3.1
After a Quench
Quench while in persistent mode:
1.
Wait until helium stops evaporating and the quench valves are closed.
2.
Wait until there is no helium vapor visible anywhere to make sure there is sufficient
oxygen in the room.
3.
Check that the globes in the quench valves are in the correct position.
4.
Remove probe and shim system to prevent formation of ice on the shim system.
5.
Start the refill with liquid helium as soon as possible.
6.
Contact Bruker Service immediately.
Figure 6.1: Magnet system during a quench
A quench is the very fast de-energizing of
the magnet by loss of its superconductivity.
The stored magnetic energy is converted
into heat, which promotes rapid evaporation
of large quantities of helium.
If a quench occurs contact Bruker Service
immediately.
WARNING
Cryogenic Agents (see
Summary of Contents for 300'89 Ascend
Page 1: ...Version Innovation with Integrity 300 89 Ascend 05 User Manual NMR...
Page 6: ...6 ZTKS0168 Z31960 05 Table of Contents...
Page 8: ...8 ZTKS0168 Z31960 05 Contact...
Page 28: ...28 ZTKS0168 Z31960 05 Safety...
Page 34: ...34 ZTKS0168 Z31960 05 Transportation...
Page 35: ...35 ZTKS0168 Z31960 05 4 Assembling Approved Persons Bruker Service only...
Page 36: ...36 ZTKS0168 Z31960 05 Assembling...
Page 57: ...57 ZTKS0168 Z31960 05 8 Disassembling Approved Persons Bruker Service only...
Page 58: ...58 ZTKS0168 Z31960 05 Disassembling...
Page 90: ...90 ZTKS0168 Z31960 05 Technical Data MS 300 89 Ascend...
Page 92: ...92 ZTKS0168 Z31960 05 Warning Signs...
Page 94: ...94 ZTKS0168 Z31960 05 List of Figures...
Page 96: ...96 ZTKS0168 Z31960 05 List of Tables...
Page 98: ...98 ZTKS0168 Z31960 05 Index...
Page 100: ...Bruker Corporation info bruker com www bruker com Order No Z31960...