21
ZTKS0168 / Z31960 / 05
Safety
Risk of Tilting
Heavy Weights
WARNING
Risk of injury due to tilting of the magnet system.
The magnet system is very sensitive to lateral forces. It may tilt.
Do not climb onto the magnet system.
Do not lean items against the magnet system.
Do not lean against the magnet system.
Do not move the magnet system on your own.
WARNING
Risk of damage to life and limb caused by heavy weights.
Lifting heavy weights is life-threatening due to falling or moving parts.
Do not stay or work under a lifted magnet system.
All lifting equipment in use must be approved to carry the weight.
Do not use damaged lifting equipment.
Do not use lifting equipment without updated check tag.
Lifting only with approved qualification.
Obey ergonomic guidelines while lifting heavy parts.
Protect parts against falling.
Always wear safety shoes with approved toe caps.
Summary of Contents for 300'89 Ascend
Page 1: ...Version Innovation with Integrity 300 89 Ascend 05 User Manual NMR...
Page 6: ...6 ZTKS0168 Z31960 05 Table of Contents...
Page 8: ...8 ZTKS0168 Z31960 05 Contact...
Page 28: ...28 ZTKS0168 Z31960 05 Safety...
Page 34: ...34 ZTKS0168 Z31960 05 Transportation...
Page 35: ...35 ZTKS0168 Z31960 05 4 Assembling Approved Persons Bruker Service only...
Page 36: ...36 ZTKS0168 Z31960 05 Assembling...
Page 57: ...57 ZTKS0168 Z31960 05 8 Disassembling Approved Persons Bruker Service only...
Page 58: ...58 ZTKS0168 Z31960 05 Disassembling...
Page 90: ...90 ZTKS0168 Z31960 05 Technical Data MS 300 89 Ascend...
Page 92: ...92 ZTKS0168 Z31960 05 Warning Signs...
Page 94: ...94 ZTKS0168 Z31960 05 List of Figures...
Page 96: ...96 ZTKS0168 Z31960 05 List of Tables...
Page 98: ...98 ZTKS0168 Z31960 05 Index...
Page 100: ...Bruker Corporation info bruker com www bruker com Order No Z31960...