Basic Setup Procedures
User Manual Version 002
BRUKER BIOSPIN
73 (327)
Calibrating 13C Pulses on Adamantane and Shimming the Probe
4.4
A high power decoupling experiment on
13
C of adamantane is used to measure
13
C pulse parameters.
NOTE: For experiments where long decoupling pulses on protons are executed,
the proton preamplifier must be bypassed, i.e. the transmitter should be wired to
the probe directly (via the proton bandpass filter) without going through the pre-
amp if a high power proton preamplifier is not available. For HP-HPPR modules
1
H/
19
F this is not absolutely necessary, but recommended. For HPLNA
1
H mod-
ules it is not required to bypass. Note that when bypassing the preamp which at-
tenuates by about 1 dB, the proton power levels should be corrected by adding 1
dB to the pl-values.
Type
edasp
in the command line. You should get a display like in
Click on
SwitchF1/F2
to
set for
13
C observation with proton decoupling. Load the
pulse program
hpdec
. Set
cpdprg2
=cw. Set
pl12
to the power level that yields a
4.5 µsec proton pulse. Set
pl1
such that in
ased
the power displayed is 200W for
13
C (7mm probe), 150W (4mm probe) or 80W (2.5 mm probe). If the green dot is
not visible in
edasp
for the
13
C channel, set pl1 to 12 dB (1 kW transmitter), 9 dB
(500W transmitter) or 7 dB (300W transmitter) for any probe. Make sure the pro-
ton channel is tuned (
wobb high
) and the carbon channel is also tuned (
wobb
).
With
d1
= 4s,
rg
= 256
swh
= 100000,
td
= 4k,
o2
set to be on resonance on the
adamantane protons as found above, accumulate 4 scans. Set the carrier fre-
quency between both adamantane
13
C peaks. Reduce the spectral width swh to
50 kHz, set
aq
=50 msec.
Acquire 2-4 scans and define the plot limits (as shown in
) for the
larger of the two peaks. Define the plot limits and determine the 90 degree carbon
pulse
p1
, using
popt
. Recalculate
pl1
for a 4.5 µsec carbon pulse using
calcpow-
lev
.
Pulse continuously using
gs
and shim the z gradient for highest FID integral.
The gradient settings can be conveniently changed in the
setsh
display.
show the adamantane
13
C FID without shims, with z-shim
adjusted and the corresponding
setsh
displays.
Summary of Contents for Solid State NMR
Page 8: ...8 BRUKER BIOSPIN User Manual Version 002 Contents...
Page 14: ...14 327 BRUKER BIOSPIN User Manual Version 002 Test Samples...
Page 118: ...118 327 BRUKER BIOSPIN User Manual Version 002 Basic CP MAS Experiments...
Page 122: ...122 327 BRUKER BIOSPIN User Manual Version 002 FSLG HETCOR Figure 8 3 The ased Display...
Page 154: ...154 327 BRUKER BIOSPIN User Manual Version 002 Proton Driven Spin Diffusion PDSD...
Page 178: ...178 327 BRUKER BIOSPIN User Manual Version 002 SUPER...
Page 192: ...192 327 BRUKER BIOSPIN User Manual Version 002 Symmetry Based Recoupling...
Page 212: ...212 327 BRUKER BIOSPIN User Manual Version 002 Relaxation Measurements...
Page 270: ...270 327 BRUKER BIOSPIN User Manual Version 002 Double CP...
Page 290: ...290 327 BRUKER BIOSPIN User Manual Version 002 Modified W PMLG...
Page 302: ...302 327 BRUKER BIOSPIN User Manual Version 002 CRAMPS 2D...
Page 304: ...304 327 BRUKER BIOSPIN User Manual Version 002...
Page 305: ......
Page 306: ......
Page 307: ......
Page 308: ......
Page 318: ...318 327 BRUKER BIOSPIN User Manual Version 002 Tables...
Page 326: ...326 327 BRUKER BIOSPIN User Manual Version 002 Index...
Page 327: ...User Manual Version 002 BRUKER BIOSPIN 327 327 End of Document...