Figure 3.37. Without/with
Parallel Capacitance to Shift the Tuning Range to Lower
Frequency
Adding a parallel capacitance does not decrease the efficiency of a circuit. How-
ever, a certain circuit has the highest possible efficiency if it is tuned with maxi-
mum inductance and minimum capacitance. Maximum inductance can usually not
be achieved due to spacial restrictions in the stator (MAS probes) or due to losses
in Q if the coil becomes too large (increase in resistance). Furthermore, capaci-
tances are tunable, inductances usually are not, so a wide tuning range can only
be achieved via exchangeable inductances and/or capacitance with a wide tuning
range. Capacitances may look different than the one shown in the picture. Fre-
quently, two larger capacitances are used in series to form a smaller capacitance
withstanding higher voltage. A parallel capacitance lowers and narrows the tuning
range.
1. X-tuning variable capacitance
2. Y-tuning capacitance
3. Proton reject filter
4. Low range extension with parallel capacitance
5. Frame flush air outlet (to protect sensitive ca-
pacitors)
Summary of Contents for Solid State NMR
Page 8: ...8 BRUKER BIOSPIN User Manual Version 002 Contents...
Page 14: ...14 327 BRUKER BIOSPIN User Manual Version 002 Test Samples...
Page 118: ...118 327 BRUKER BIOSPIN User Manual Version 002 Basic CP MAS Experiments...
Page 122: ...122 327 BRUKER BIOSPIN User Manual Version 002 FSLG HETCOR Figure 8 3 The ased Display...
Page 154: ...154 327 BRUKER BIOSPIN User Manual Version 002 Proton Driven Spin Diffusion PDSD...
Page 178: ...178 327 BRUKER BIOSPIN User Manual Version 002 SUPER...
Page 192: ...192 327 BRUKER BIOSPIN User Manual Version 002 Symmetry Based Recoupling...
Page 212: ...212 327 BRUKER BIOSPIN User Manual Version 002 Relaxation Measurements...
Page 270: ...270 327 BRUKER BIOSPIN User Manual Version 002 Double CP...
Page 290: ...290 327 BRUKER BIOSPIN User Manual Version 002 Modified W PMLG...
Page 302: ...302 327 BRUKER BIOSPIN User Manual Version 002 CRAMPS 2D...
Page 304: ...304 327 BRUKER BIOSPIN User Manual Version 002...
Page 305: ......
Page 306: ......
Page 307: ......
Page 308: ......
Page 318: ...318 327 BRUKER BIOSPIN User Manual Version 002 Tables...
Page 326: ...326 327 BRUKER BIOSPIN User Manual Version 002 Index...
Page 327: ...User Manual Version 002 BRUKER BIOSPIN 327 327 End of Document...