STMAS
User Manual Version 002
BRUKER BIOSPIN
251 (327)
because it is essential that the centres of the pulses
P1
and
P2
are exactly an in-
teger number of rotor periods apart. In this formula
P1
,
P2
, and
P4
are the RF
pulses as listed in table 2,
CNST31
must be set equal to the spinning frequency.
This means that the first increment can last between 100 µs (10 kHz spinning)
and 28.5 µs (35 kHz spinning). Since
P4
is the 180° selective pulse, which can be
as long as 40 µs,
L0
must be set large enough to avoid the situation where the
calculated
d0
is negative. Optimization of the pulses
P1
and
P2
can be done us-
ing
popt
in full analogy to the optimization of the pulses in MQMAS.
Two Dimensional Data Acquisition
19.3.2
Once the pulses are calibrated the 2D data acquisition can be used to find the cor-
rect and precise magic angle setting. Create a new data set and change
parmode
to 2D. The acquisition parameters for the (new) indirect F1 dimension must be set
according to
.
Similar considerations for the maximum t
1
period, determined by the number of
FID’s to be acquired and the t
1
increment, can be made as for MQMAS. Because
the shift range in ppm is twice a big as in 3QMAS a larger increment can be used
to give an equivalent shift range, the increment being calculated from the spinning
speed. Since the magic angle is probably not yet perfect 32 to 64 FID’s will be suf-
ficient initially. Processing parameters are described in the next section.
Table 19.4.
Initial Parameters for the Set-up of stmasdfqe.av.
Parameter
Value
Comments
pulprog
stmasdqfe.av
Pulse program.
NS
16*n
For set-up the full phase cycle is not so critical.
D0
See text
Calculated in pulse program.
D1
5 * T
1
Recycle delay, use dummy scans if shorter.
D6
See text
D7
See text
P1
1.5 µs
Excitation pulse at pl11.
P2
1.5 µs
Conversion pulse at pl11.
P3
20 µs
90° selective pulse at pl21 taken from previous pulse cali-
bration.
PL1
=120 dB
Not used.
PL11
Start with
≈
150 to 300 W
Power level for excitation and conversion pulses.
PL21
Power level for selective pulse, approx. pl11+30 dB taken
from previous pulse calibration.
Summary of Contents for Solid State NMR
Page 8: ...8 BRUKER BIOSPIN User Manual Version 002 Contents...
Page 14: ...14 327 BRUKER BIOSPIN User Manual Version 002 Test Samples...
Page 118: ...118 327 BRUKER BIOSPIN User Manual Version 002 Basic CP MAS Experiments...
Page 122: ...122 327 BRUKER BIOSPIN User Manual Version 002 FSLG HETCOR Figure 8 3 The ased Display...
Page 154: ...154 327 BRUKER BIOSPIN User Manual Version 002 Proton Driven Spin Diffusion PDSD...
Page 178: ...178 327 BRUKER BIOSPIN User Manual Version 002 SUPER...
Page 192: ...192 327 BRUKER BIOSPIN User Manual Version 002 Symmetry Based Recoupling...
Page 212: ...212 327 BRUKER BIOSPIN User Manual Version 002 Relaxation Measurements...
Page 270: ...270 327 BRUKER BIOSPIN User Manual Version 002 Double CP...
Page 290: ...290 327 BRUKER BIOSPIN User Manual Version 002 Modified W PMLG...
Page 302: ...302 327 BRUKER BIOSPIN User Manual Version 002 CRAMPS 2D...
Page 304: ...304 327 BRUKER BIOSPIN User Manual Version 002...
Page 305: ......
Page 306: ......
Page 307: ......
Page 308: ......
Page 318: ...318 327 BRUKER BIOSPIN User Manual Version 002 Tables...
Page 326: ...326 327 BRUKER BIOSPIN User Manual Version 002 Index...
Page 327: ...User Manual Version 002 BRUKER BIOSPIN 327 327 End of Document...