Relaxation Measurements
User Manual Version 002
BRUKER BIOSPIN
209 (327)
Data Processing
The saturation-recovery data should be processed in the same way as the inver-
sion-recovery data above (see
for parameters). The only differences
are that the fitting function should be satrec rather than expdec, and the slice se-
lected for processing should be the last one (signal is maximum at long recovery
times). The calculated relaxation time constants should be the same as those ob-
tained by inversion-recovery.
T1p Relaxation Measurements
16.2.5
Rotating-frame relaxation measurements, under a spin-locking rf field, can be
used to probe motions on shorter domestically than T
1
measurements, with in-
verse correlation times of the order of the spin-locking rf field strength.
To measure T
1r
relaxation, after CP a variable length spin-locking pulse is applied
to the X nucleus. The remaining X magnetization decays exponentially to zero, as
a function of spin-lock time. The parameters of cross-polarization can also be de-
termined from variable-contact-time CP experiments (the function cpt1rho is pro-
vided in the relaxation analysis tool for this purpose), but here only simple T
1r
measurements will be discussed.
It should be noted that the relaxation in a T
1r
experiment might result from pro-
cesses other than true T
1r
relaxation. For example, in glycine, the carbon spins
are dipolar coupled to protons, and there is a possible fast relaxation pathway via
the protons, which is not T
1r
relaxation. This is inhibited by having a high spin-lock
field strength, but at large field strengths care must be taken over the length of the
spin-lock pulse. If apparently non exponential decay is observed, this may result
from such alternative relaxation processes.
Experiment setup
Sample
: Glycine
Spinning speed
: 10 kHz
Time
: 20 minutes
Table 16.4.
Parameters for the Saturation Recovery Experiment
Parameter
Value
Comments
Parmode
2D
Vdlist
See text
td(f1)
Number of entries in vd list
FnMODE
QF
This is not a real 2D experiment.
NS
16, for the glycine sample
More scans needed than for the CP experiment,
due to reduced signal.
Summary of Contents for Solid State NMR
Page 8: ...8 BRUKER BIOSPIN User Manual Version 002 Contents...
Page 14: ...14 327 BRUKER BIOSPIN User Manual Version 002 Test Samples...
Page 118: ...118 327 BRUKER BIOSPIN User Manual Version 002 Basic CP MAS Experiments...
Page 122: ...122 327 BRUKER BIOSPIN User Manual Version 002 FSLG HETCOR Figure 8 3 The ased Display...
Page 154: ...154 327 BRUKER BIOSPIN User Manual Version 002 Proton Driven Spin Diffusion PDSD...
Page 178: ...178 327 BRUKER BIOSPIN User Manual Version 002 SUPER...
Page 192: ...192 327 BRUKER BIOSPIN User Manual Version 002 Symmetry Based Recoupling...
Page 212: ...212 327 BRUKER BIOSPIN User Manual Version 002 Relaxation Measurements...
Page 270: ...270 327 BRUKER BIOSPIN User Manual Version 002 Double CP...
Page 290: ...290 327 BRUKER BIOSPIN User Manual Version 002 Modified W PMLG...
Page 302: ...302 327 BRUKER BIOSPIN User Manual Version 002 CRAMPS 2D...
Page 304: ...304 327 BRUKER BIOSPIN User Manual Version 002...
Page 305: ......
Page 306: ......
Page 307: ......
Page 308: ......
Page 318: ...318 327 BRUKER BIOSPIN User Manual Version 002 Tables...
Page 326: ...326 327 BRUKER BIOSPIN User Manual Version 002 Index...
Page 327: ...User Manual Version 002 BRUKER BIOSPIN 327 327 End of Document...