Technical Manual Version 002
BRUKER BIOSPIN
47 (55)
5
Technical
specifications
5
Specifications
5.1
Table 5.1.
BVT3200 Specifications
Temperature controller
Eurotherm Model 2416 (display units K or °C).
Sensors
T thermocouple with compensated connector.
BTO2000 or PT100.
Temperature range
35°C to 200°C without special accessory.
•
Low temperature limit :
- 5° C with BCU05 cooler.
-150 °C with LN
2
exchanger or LN
2
evaporator.
•
High temperature limit :
600 °C with power booster BVTB3500
Temperature stability
0.01 °C/°C for sample temperature between 0 °C and 50 °C, using a
BTO2000 and BCU05.
0.1 °C/°C without BTO2000.
Heater power
135 W (48 V max. on 12
Ω
probe heater).
Output current limitation at 6A max.
Gas inlet
Dry air or nitrogen 4 bar mini and 7.5 bar maxi.
Dry air, dew point < 5 °C for operation above room temperature.
Dry air, dew point < -50 °C for operation with BCU05 cooler.
Gas flow rate
200 l/h to 2000 l/h.
Fifteen values available.
Controlled by software.
Control
Computer control via RS232 interface.
Software control
By EDTE XWINNMR 2.5 or later, TOPSPIN.
Options
LN
2
evaporator (210W, with optional printed circuit board and 400 VA power
supply).
LN
2
exchanger (with optional printed circuit board).
BCU05 gas cooler.
BVTB3500 External Power Booster.
BVT3900 Probe heat exchanger.
Summary of Contents for NMR BVT3200
Page 16: ...16 55 BRUKER BIOSPIN Technical Manual Version 002...
Page 24: ...24 55 BRUKER BIOSPIN Technical Manual Version 002...
Page 50: ...50 55 BRUKER BIOSPIN Technical Manual Version 002 Figures...
Page 52: ...52 55 BRUKER BIOSPIN Technical Manual Version 002 Tables...
Page 55: ...Technical Manual Version 002 BRUKER BIOSPIN 55 55 End of Document...