Weatherproof Microphone Unit Type 4184-A – Supplement to Instruction Manual BE 1712
18
The Lower Limit in the following sections is based on the guaranteed worst-case limit for the analyzer and
the nominal Open Circuit Sensitivity of the microphone, under Reference Environmental Conditions,
Sound
Field Correction
set to
Free-field
and no microphone accessories selected.
The maximum Sound Level that the Sound Level Meter can accommodate without causing damage to the
Sound Level Meter: 176 dB Peak.
Total Range is defined as the difference between the Upper Limit on the least sensitive level range, and the
lowest sound pressure level measurable on the most sensitive level range, which can be measured at 1 kHz
within the most conservative tolerance limits, specified in the International Standards IEC 61672–1,
IEC 60651 and IEC 60804:
NOTE:
For Sound Exposure Levels, the stated ranges are valid if 10*lg(
t
) is added to the limits.
t
being
the averaging time interval, indicated as
Elapsed Time
, expressed in seconds.
Primary Indicator Range according to the International Standard IEC 60651:
Table 4.3
Total range
Frequency Weighting
A-weighting
(dB)
B-weighting
(dB)
C-weighting
(dB)
Z-weighting
(dB)
Z-weighting
Extended Low
Frequency
(dB)
150.3 – 40.2
150.3 – 38.8
150.3 – 39.3
150.3 – 44.5
150.3 – 55.5
Table 4.4
Primary Indicator
Range
Range
Upper
Limit
(dB)
Lower Limit
A-weighting
(dB)
B-weighting
(dB)
C-weighting
(dB)
Z-weighting
(dB)
Z-weighting
Extended Low
Frequency
(dB)
Single
132.3
38.9
37.5
38.0
43.2
54.2
High
132.3
57.8
56.2
56.2
61.0
61.8
Low
105.8
38.9
37.5
38.0
43.2
54.2
Summary of Contents for 4184-A
Page 2: ......
Page 6: ......
Page 42: ...Weatherproof Microphone Unit Type 4184 A Supplement to Instruction Manual BE 1712 36...
Page 48: ...Weatherproof Microphone Unit Type 4184 A Supplement to Instruction Manual BE 1712 42...
Page 50: ...Weatherproof Microphone Unit Type 4184 A Supplement to Instruction Manual BE 1712 44...
Page 51: ......