SETUP - Adjustments
VIBROTEST 60
Page 46 of 168
C102 853.002
© VTEST60E July 2013 4.46ff
4.5.2 Envelope analysis - SED mode (
S
elective
E
nvelope
D
etection)
OK The displayed parameter setup resp.
adjustments are accepted and the previous
menu appears.
EDIT Switch to the functional selection/
entry menu for editing the parameter
values.
4.5.2.1 Sensors
AS-06x/07x
Vibration acceleration sensor
AS-1x/2x/3x
Vibration acceleration sensor
(only with additional adapter AC-630)
Variable
Sensor not requiring power from the
VIBROTEST 60
0.01 ... 1000.00
mV / unit
V / unit
Variable CCS*
Sensor requiring power from the
VIBROTEST 60
+ 24 V, 2 ...4 mA
0.01 ... 1000.00
mV / unit
V / unit
4.5.2.2 Unit / Signal detection
Unit
Signal detection
AS - Sensors:
g
rms
m/s²
peak
pp
peak peak
Variable Sensors:
Sensor-definition unit
rms
peak
pp
peak peak