High Voltage
High voltage electric shock could cause severe injury or
loss of life.
Equipment Failure
To avoid equipment failure, use appropriate guage wire per local electrical
codes and 3 conductor cable with ground rating at 600 VAC for the
Compressor.
High Capacity Helium Compressor
Installation
Manual
Electrical Connections
Brooks Automation
8040600
Revision A
4-5
Electrical Connections
The following procedures provides information for making the input power
connections to the High Capacity Helium Compressor.
Power Cable Preparation
1. Remove the electrical panel cover by turning the four captive 1/4
turn screws as shown in
Figure 4-2
.
2. Remove the terminal block cover by turning the two 1/4 turn screws.
The conductor isolator blocks will be inside the terminal block cover as
shown in
Figure 4-2
.