5.3
Common issues
Symptom
Possible cause
Action
No (fieldbus) communication
No power supply
·
Check power supply
·
Check cable connection
·
Check cable hook-up
Invalid node address
Change node address (see
Other
Reset instrument and/or restart master. If
problem persists, contact Bronkhorst.
No output signal
No power supply
·
Check power supply
·
Check cable connection
·
Check cable hook-up
Invalid control mode (instrument
accepts no setpoint)
Check control mode (see
No setpoint given or setpoint too
low
Give setpoint 2%
Control valve in Safe State
(normally closed)
Check if control valve is in safe state; solve
cause if necessary (see
Inlet pressure or differential
pressure too low
Increase inlet pressure
Piping, filters and/or control valve
clogged or blocked
·
Flush fluid system with clean, dry air. If
problem persists, contact Bronkhorst.
·
For external proportional control valves:
supply 0…15 Vdc and operational inlet
pressure to valve and slowly increase
voltage. If valve does not open, clean parts
and re-adjust valve
Laminar flow element clogged or
blocked
Return equipment to factory
Sensor failure
Return equipment to factory
Control behavior unstable
Measurement disturbed by
vibrations
If possible, avoid installation in close proximity
of mechanical vibration
Inlet pressure unstable
Install pressure regulator or increase buffer
volume between controlling instruments (see
section
Inlet and/or outlet pressure too
high or too low
Adjust pressure and/or set instrument pressure
in accordance with actual process pressure (e.g.
with FlowTune™)
Wrong process gas selected
Select correct process gas (e.g. with FlowTune™)
Wrong controller settings
Adjust settings (e.g. with FlowPlot)
Control valve damaged
Return equipment to factory
No flow (sending a setpoint has no
effect)
No fluid supply
Check upstream components for obstruction,
e.g.:
·
fluid lines
·
valves
·
filters
Setpoint too low
Give setpoint 2%
Inlet pressure or differential
pressure out of bounds
Set inlet pressure to a value within
specifications