Quick Installation Guide
Doc. no.: 9.17.098 rev. C Date: 06-04-2020
Starting up EL-FLOW®
Select
in 10 steps
Page 1: ...Quick Installation Guide Doc no 9 17 098 rev C Date 06 04 2020 Starting up EL FLOW Select in10 steps ...
Page 2: ...Manual EL FLOW Select series document no 9 17 099 Interfaces Manual Hook up diagram Analog RS232 interface 9 17 027 9 16 119 EtherNet IP interface 9 17 132 9 16 215 CANopen interface 9 17 131 9 16 217 DeviceNet interface 9 17 026 9 16 122 EtherCAT interface 9 17 063 9 16 124 FLOW BUS interface 9 17 024 9 16 120 Modbus ASCII RTU interface 9 17 035 9 16 123 PROFIBUS DP interface 9 17 025 9 16 121 PR...
Page 3: ...e flow arrow on the instrument body Tighten the fluid connections according to the instructions of the supplier of the fittings The use of Swagelok RS type stainless steel adapters is recommended Mounting the instrument in the upright position is preferred especially if the operating pressure will be higher than 10 bar When mounting the instrument in a position with upward or downward flow adjusti...
Page 4: ...tem to air Warm up time For accurate results it is recommended to warm up the instrument for at least 30 minutes before starting measurement this can be done while purging 10 Adjusting zero point The zero point of each Bronkhorst flow meter is factory adjusted If required the zero point can be re adjusted via RS232 or with the multifunctional switch Procedure for adjustment with the multifunctiona...