9
Figure 10 - Current Adjustment 0A
To re-calibrate briefly press the Command Button
and the following information will be
displayed:
Figura 11 – Ajuste de corrente STD (Standard)
The STD current must equal the production capacity of the installed cell (8G = 8A; 10G = 10A,
etc ...). If the current is higher or lower, you must turn off the equipment not to save the calibration
and contact the manufacturer's technical service or your local representative.
6.2
ORP Calibration
The calibration of the ORP probe is done with 2 buffers (240mV and 470mV). Before
proceeding, confirm that you have all the necessary material including a glass of water and
absorbent paper in order to clean the probe before introducing the buffers.
To enter the ORP calibration, select the C.O option in the Calibration / Configuration menu (see
chapter 6).
When entering calibration the following information will appear on the screen:
Figure 12 - Inserting the ORP probe into the standard 240mV
Insert the probe into the 240mVbuffer, wait 10 seconds, and briefly press the Command
Button
. The first line of the display indicates a default value, and the second line shows the
actual reading and a dashed line indicating that the system is calibrating (Fig. 13).
Figure 13 - Standard ORP Calibration 240mV
When the calibration is completed at 240mV, "OK" appears on the display and the buffer
470mV is requested. Repeat the entire process with the 470mV buffer until the calibration is
complete.