AWT35-100089 Rev AA
19/25
4.3.2 Calibration Procedure
The procedure covers the
Line Calibration
. You can add up to 1+3 calibration points.
Calibration and/or configuration of calibration settings of your scale should be
accomplished by a trained service technician using certified weights to ensure
proper operation and accuracy. Calibration is not covered under warranty.
1. Turn the scale on and zero the display, if necessary, by pressing the
ZERO/ON/OFF
key.
2. Press the
TARE
and
ZERO/ON/OFF
keys. Release the keys when the “CAL” message pops up
to open the Calibration Menu. The number of calibrations is shown.
3. Press the
BMI
key to select the “LINE” menu option.
4. Make sure that there is no weight on the platform scale.
5. Press the
TARE
key to open the
Line Calibration
mode.
CAL.P0, and 0 will be displayed. Press the
UNIT
key for more than 2.5s to change the
calibration unit.
6. Press the
TARE
key to store the Zero Calibration Point.
CAL.P1, and the scale capacity will be displayed.
8. Place the certified weight over the scale platform.
9. Press the
TARE
key to store the CAL.P1 Point.
10. Press the
TARE
key to save the calibration points and exit the Calibration mode or, press the
key to add CAL.P2/3 points.
On calibration complete, the indicator will reboot and return to general weighing mode.
If there’s an error occurred in calibration, the message “CAL.Er” is displayed. Repeat steps 5 to 9 to
adjust the calibration points.
4.4 Service Configuration Menus
The scale is configured from the factory to match the specified settings for each unit, as defined by the
product specifications and sales brochure. Modification of the SBI 210-LCD Indicator settings can be
accomplished by altering the configuration settings in the Config Menu.
The configuration and test of your scale should be accomplished by a trained
service technician.
4.4.1 Configuration Options
In power-off mode, press the
HOLD
and
ZERO/ON/OFF
keys.
Release the
ZERO/ON/OFF
key when the display lights.
Release the
HOLD
key when the message “CONFIG” pops up to open the Configuration Menu.
The number of time that the indicator has been configured is displayed.
Navigate the menu using the keys, press the
TARE
key to confirm or the
ZERO/ON/OFF
key to cancel the operation / exit the menu.