Replacing the Patient Circuit

Replacing the Patient Air Inlet Filters

Stop treatment before disconnecting the 
patient circuit.

Connecting a single limb patient circuit:

1. Connect the patient circuit to the patient air outlet 

of the ventilator.

2. Connect the control pressure tubing to the exhalation 

valve control pressure outlet of the Vivo 50.

3. Connect the other end of the patient circuit to the 

patient interface.

Symbols

SYMBOL DESCRIPTION

Home Mode activated

Leakage circuit selected (single limb)

Exhalation valve circuit selected (single limb)

iOxy connected

FiO

2

 connected

CO

2

 connected

Multiple pages

Multiple content available

High priority alarm event in history list

Medium priority alarm event in history list

The filters are located in 
the filter cassette at the 
side of the ventilator. 

Washable Filter

 (grey): 

Replace at least once a 
year, wash at least once a 
week.

Disposable Filter

 (white): 

Using Oxygen

• O

2

 must be turned off when the  

Vivo 50 is not in operating mode!

• It is recommended to monitor the  

oxygen concentrations.

• Oxygen flow must not exceed 15 l/min.

1. Mount the oxygen tube 

to the connector.

2. Connect the oxygen 

tube. A click sound is 
heard when the tube is 
properly attached.

Disconnect:

1. Press the release  

button to disconnect 
the oxygen tube.

2. Remove the tube.

3. Turn off the oxygen 

source.

Using the Connectors

Make sure that the color 
mark is facing up.

Pull the connector sleeve, 
not the cable itself to 
release the connector.

Replace at least every 4th week, or more often in high 
pollution or pollen-rich environments.