7.4.5
Ambient Pressure Compensation Lost Alarm
Property
Description
Alarm text
Pressure Comp Lost
Priority
Medium
Alarm condition
An Ambient Pressure Compensation Lost alarm will be given
when the automatic ambient pressure compensation func-
tionality is out of order.
Ventilator action
The Vivo 3 will continue treatment according to the current
settings.
Normal atmospheric pressure at sea level will be used as
approximation for the temporary ambient pressure compen-
sation. If used at other altitude, delivered and measured pres-
sures may deviate.
Reset
Reset of ventilator.
7.4.6
Low Battery Alarm
Property
Description
Alarm text
Low Battery. Connect to Mains.
Priority
Medium
Alarm condition
This alarm will be given when the last battery source (internal
battery) has 15 minutes of operating time left with current
settings.
Ventilator action
The Vivo 3 will continue treatment according to the current
settings.
Reset Criteria
Connection of another power source within 15 minutes or
the treatment will shut down.
7.4.7
Critically Low Battery Alarm
Property
Description
Alarm text
Critically Low Battery. Connect to Mains
Priority
High
Alarm condition
This alarm will be given when the last battery source (internal
battery) has 5 minutes of operating time left with current
settings.
Ventilator action
The Vivo 3 will continue treatment according to the current
settings.
Reset Criteria
Connection of another power source within 5 minutes or the
treatment will shut down.
106 Alarms
Vivo 3 Clinician
’
s Manual
Doc. 007232 Y-2