REMOVING/REPLACING THE MAIN COMPONENTS
PV 101+/PV 102+ Service Manual
3–2
BREAS MEDICAL
Doc. No. 003259 En
Issue: X-1
3.2
Removing and reassembling the upper casing
Disconnect the power supply before removing the casing of the machine.
To remove the casing:
1
Remove the mains power cord.
2
Turn the PV 101+/PV 102+ upside down. Remove the four screws placed in the
four holes underneath the lower casing and the two screws placed underneath the
handle.
3
Lift up the upper casing. Disconnect the CPU board’s ribbon cable from the MDA
board. Put the upper casing aside.
To reassemble the casing:
•
Reassemble in reverse order.
6
×
TX25