PV 101+/PV 102+ Service Manual
MAINTENANCE SERVICE INSTRUCTIONS
2.4
External inspection
2.4.1
Visual inspection for external damage and wear
1
Clean the cover of the PV 101+/PV 102+ using a mild detergent.
2
Check for visible damage to the cover and other components.
3
Check that all external parts are tightened and secured.
2.4.2
Checking the power supply connection
1
Check the plug, power cord and mains power socket.
2
Make sure the strain-relief clamp for the power cord is not damaged.
3
Check the power supply to the machine.
4
If the PV 101+ does not operate from the mains power, check the fuses.
2.4.3
Short function check
1
Connect the power cord to the power socket of the PV 101+/PV 102+.
2
Connect the patient circuit.
3
Start the PV 101+/PV 102+ and check that it functions normally.
4
Disconnect the power cord and check that the power failure alarm sounds. Recon-
nect the power cord. The alarm should be silenced.
5
Switch on the PV 101+/PV 102+. Upon starting, all the LEDs on the front panel
should be lit and the trigger settings should be indicated on the bar graph. The
entire bar graph should then become yellow for 2 seconds before the device starts
functioning normally.
6
(PV 102+ only) Switch on the PV 102+ and wait for 15 seconds for the leakage
alarm to be activated.
7
Switch off the device.
Summary of Contents for PV 101+
Page 76: ...FAULT TRACING PV 101 PV 102 Service Manual 8 2 BREAS MEDICAL Doc No 003259 En Issue X 1...
Page 78: ...9 2 BREAS MEDICAL Doc No 003259 En Issue X 1 APPENDICES PV 101 PV 102 Service Manual...
Page 82: ...9 6 BREAS MEDICAL Doc No 003259 En Issue X 1 APPENDICES PV 101 PV 102 Service Manual...
Page 87: ...BREAS MEDICAL 10 PV 101 OPERATING MANUAL...
Page 88: ...BREAS MEDICAL...
Page 89: ...BREAS MEDICAL 11 PV 102 OPERATING MANUAL...
Page 90: ...BREAS MEDICAL...