4
Pull out the board
’
s flow sensor
out of its attachment.
5.8
Replace the CPU Board
1
Remove the top cover according to 5.19
2
Remove the front cover according to 5.17
The front, including the CPU board, is now separated from the ventilator.
3
Remove the four screws.
Torque: 0.7 Nm
Size: M3x6
4
Disconnect the cable between the
CPU board and the display.
5.8.1
Replace the Humidifier Sense Board
1
Remove the top cover according to 5.19
2
Remove the front cover according to 5.17
The front, including the CPU board, is now separated from the ventilator.
48 Opening the Ventilator and Replacing the Main Components
Vivo 45/Nippy 4 Series Service manual
Doc. 007132 D-2