2.9.1
How to perform the CO
2
Gas Span Check
1
Snap on a new airway adapter to
the CO
2
sensor probe.
2
Perform a CO
2
zeroing procedure.
You will find the zeroing procedure under
“
FiO2/CO2 Calibration
”
in the
“
Others
”
section
3
Mount the gas regulator on the gas
bottle.
Be careful not to damage the thread
on the bottle.
Make sure that the pressure
indicator is not within the red mark.
Doc. 007132 D-2
Vivo 45/Nippy 4 Series Service manual