Alarms
Vivo 45 User Manual
Doc.006149 En Y-2
Low FiO
2
Alarm
I
TEM
D
ESCRIPTION
Definition
A Low FiO
2
alarm will be given when the meas-
ured FiO
2
is below the alarm limit for
30 ±0.5 seconds.
Priority
High
Possible cause
• Decreased oxygen inlet.
• Disconnection of oxygen inlet.
• Increased minute ventilation.
• High leakage.
Setting min
Off, 21%
Setting max
100%
Setting resolution
1%
Ventilator action
The Vivo 45 will continue treatment with the
same settings.
Indication