background image

Probe Covers 

Probe covers 

 

Are intended for 

single use only

 

Always use PRO 4000 disposable probe covers. 

The new PRO 4000 probe covers can be used with all 
models of Braun ear thermometers.   

Older version probe covers (i.e. those purchased for the 
PRO 3000) cannot be used with the PRO 4000.  The 
result will be inaccurate readings. 
 

Damaged, perforated, or soiled covers can lead to 
inaccurate temperatures. 
 

The device will not take a temperature without a probe 
cover in place. 

MC12687 

Summary of Contents for WelchAllyn ThermoScan PRO 4000

Page 1: ...Welch Allyn Braun ThermoScan PRO 4000 Inservice MC12687 1...

Page 2: ...ning the Electrical Contacts 4 Probe Covers 5 Tympanic Measurements 6 Interferences 7 Contraindications 8 Error Messages and Solutions 9 Base Station and Anti Theft Function 10 Memory Mode 11 Temperat...

Page 3: ...accurate and faster readings ExacTempTM light to ensure proper steady positioning Safe Probe cover eject button designed to reduce the risk of cross contamination Secure Optional security features can...

Page 4: ...minutes drying time before taking temperatures Avoid touching the window except when cleaning required Only alcohol should be used to clean the window Thermometer Body Attach a probe cover to protect...

Page 5: ...ntacts regularly Every 3 months OR Any time the charging or communication performance between the thermometer and dock becomes erratic Ex Incomplete charging or thermometer fails to transfer a reading...

Page 6: ...be used with all models of Braun ear thermometers Older version probe covers i e those purchased for the PRO 3000 cannot be used with the PRO 4000 The result will be inaccurate readings Damaged perfor...

Page 7: ...Tympanic Measurements Proper Probe Placement Guide MC12687 7...

Page 8: ...from the situation and wait 20 minutes prior to taking a temperature DOES affect temperature accuracy A used probe cover Wet dirty damaged probe lens Hearing aid Lying on a pillow DOES NOT affect temp...

Page 9: ...Not intended for pre term babies or small for gestational age babies Complete ear canal occlusion due to cerumen ear wax can result in lower temperature readings Use the untreated ear if prescription...

Page 10: ...Error Messages and Solutions MC12687 10...

Page 11: ...heft function Can be wall mounted for convenient storage Anti Theft Function Requires returning the thermometer to the station within an individually pre selected time If not returned to the station t...

Page 12: ...ure press the mem button to recall the last reading The last temperature taken before the thermometer is turned off is stored in its memory and will be automatically displayed when it is turned on aga...

Page 13: ...d off 2 Press and hold the mem button After approximately 3 seconds the display will show F SET C SET 3 Release the mem button when C is shown 4 A short beep will confirm the new setting and then it w...

Page 14: ...me with separately purchased base station Average battery life is 6 months or 1000 measurements Battery Replacement 1 Open battery compartment using a pen 2 Remove old batteries 3 Insert new batteries...

Page 15: ...Resources Thank you for choosing Welch Allyn Customer Care 1 800 535 6663 Website www welchallyn com MC12687 15...

Reviews: