Unit Elements
4
Perfusor® compact, 3.0 gb
4 - 13
this straight pin lock is already fitted. Please read instructions
provided.
5. Insert syringe size board.
6. Mount scraper ring and axial positioner and screw down
drive.
7. Snap in syringe size board on both sides.
Note
Always fasten syringe size board with screws to centering bearing.
8. Pay attention to cable laying (please see
9. Close unit
Note
Do not squeeze cable.
10. Calibrate in Service Program
4.14 Axial Positioner
Designation
Ord. No.
Axial positioner . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 6659
Exchange
1. Open unit
2. Move drive arm to middle position and lock.
3. Loosen both screws on drive.
4. Lift drive until the axial positioner is free.
5. Remove axial positioner by forcing apart. Replace new axial
positioner and make sure that the scraper ring is correctly
fitted.
6. Assembly is performed in reverse order.
Note
Do not squeeze the cable
7. Calibration is required in Service Program
, as the drive was dismounted.
Fig.: 4 - 15
3.0
Summary of Contents for perfusor compact
Page 1: ...Perfusor compact Service Manual 0 10 3 1 0 Version 3 1 English...
Page 10: ...Contact Persons 0 0 10 Perfusor compact 2 1 gb For your notes 2 1...
Page 14: ...System Overview 1 1 4 Perfusor compact 2 1 gb For your notes 2 1...
Page 20: ...Software 2 2 6 Perfusor compact 2 1 gb For your notes 2 1...
Page 56: ...Unit Elements 4 4 18 Perfusor compact 2 1 gb 2 1 For your notes...
Page 62: ...Checks after Repair 5 5 6 Perfusor compact 2 1 gb 2 1 For your notes...
Page 64: ...Maintenance 6 6 2 Perfusor compact 2 1 gb For your notes 2 1...
Page 72: ...Procedural Instructions on the TSC 8 8 6 Perfusor compact 2 1 gb 2 1 For your notes...
Page 74: ...Test Equipment and Special Tools 9 9 2 Perfusor compact 2 1 gb For your notes 2 1...
Page 78: ...Spare Parts List 10 10 4 Perfusor compact 2 1 gb For your notes 2 1...
Page 80: ...Revision Documentation 11 11 2 Perfusor compact 2 1 gb 2 1...
Page 85: ......