Turning the equipment on
Part 1 • Page 9-1
Part 1 – Operating Instructions
Rev. 6.4 dated 2022-02
Software Released Version: 1
(Eco)RO Dia II
Part 1 • Chapt
er 9
9.
Turning the equipment on
wd dd.mm.yy hh:mm
EcoRO Dia II
CPU test
Menu
Operating status and phase
To supply the system with power, insert the mains plug and turn the main
switch to ON.
wd dd.mm.yy hh:mm
System off
Menu
Switch the operating mode switch to 0 = System off.
After an initial test of the control unit, the display will show the current
operating mode “System off”.
Operating mode switch, switch position 0:
Switch position 0 means the system is off, all fault messages are reset.
Operating mode switch, switch position 1:
Switch position 1 means dialysis mode.
The remote control input and menu 3 auto on/off are not taken into
account.
Operating mode switch, switch position 2:
Switch position 2 means standby mode/automatic mode.
The remote control and menu 3 auto on/off can be used to switch between
standby mode and dialysis mode, emergency mode RO I, emergency
mode RO II.
wd dd.mm.yy hh:mm
Alarm
Menu Reset
Fault message
(wd dd.mm.yy hh:mm = weekday, day, month, year / hour, minute).
If a fault is determined in this mode of operation, the fault message
(“Alarm” or “Error”) is given in plain text.
Depending on the fault type, it can be acknowledged by pressing the
>Reset<
button.
NOTE
After the power supply has been interrupted, the system always switches to the mode of
operation that was last active.