a

 

BL-IL-60960-34EN Rev2.0 

*60960-34EN* 

 

 

Converter Kit DVI to S-Video 

19179

 

 

    

 

ENGLISH  

 

Intended Use 

The 

Converter Kit

 is intended to be used by surgeons or other 

clinical  personnel  in  OR  environment  during  image  guided 
surgery (IGS) for integration of video signals from video sources 
with  only  DVI-output  to  navigation  platforms  without  DVI-IN 
interfaces. 
 

Overview 

           

 

 

 

          

 

 

No. 

Component 

 

S-Video OUT

 

 

Mains Power Connector 

 

Ports for Video and Audio Output

 

 

HDMI Video IN

 

 

USB (for service purposes only)

 

 

Power Supply Connector

 

 

Power LED Status 

Color

 

Status

 

Explanation

 

Blue 

Off 

No power connection, electronic 
components are off 

Solid 

Connection to mains power, 
electronic components are on 

 

Compatibility 

Only use the 

Converter Kit

 with Brainlab navigation systems: 

 

Kick 

 

Curve 

 

Only use compatible C-arms. 
For a complete compatibility list, contact Brainlab support. 
Brainlab  accepts  no  liability if  unapproved  scanners  or  C-arms 
are used. 

Video Connections 

Port 

Example 

Handling 

S-Video 
OUT 

 

Connect the 

Converter Kit

 to 

the 

Brainlab 

navigation 

system with an S-Video cable 
to  interact 

with 

Brainlab 

software video. 

HDMI 
IN 

 

Connect 

the 

DVI 

video 

source  to  the 

Converter  Kit

 

with a DVI to HMDI cable. 

NOTE:  Use  a  high  quality 
HDMI  cable  to  connect  a 
HDMI source. 

 

 

Do not place the Converter Kit in the patient environment. 
 

 

Additional  equipment  connected  to  medical  electrical 
equipment  must  comply  with  the  respective  IEC  or  ISO 
standards  (e.g.  IEC  60950  for  data  processing  equipment 
and  IEC  60601-1  for  medical  equipment).  Furthermore,  all 
configurations  shall  comply  with  the  requirements  for 
medical electrical systems (see IEC 60601-1-1 or clause 16 
of  the  3rd  edition  of  IEC  60601-1,  respectively).  Anyone 
connecting  the  Converter  Kit  to  medical  electrical 
equipment  configures  a  medical  electrical  system  and  is 
therefore  responsible  that  the  system  complies  with  the 
requirements  for  medical  electrical  systems.  Please  note 
that  local  laws  take  priority  over  the  above-mentioned 
requirements. If in doubt, consult your local representative 
or the technical service department. 
 
 

NOTE:  Unless  otherwise  specified,  the  use  of  multiple  socket 

outlet(s) is not permitted.

 

 
 
 

 

 
 
 

Startup Test 

Perform a functional test before first use at each day of use. To 
do  this,  the 

Converter  Kit

  must  be  connected  to  the  video 

source and the Brainlab navigation System as described above. 
 
Verify that the image shown on the Brainlab navigation system 
is correct. 
 

Cleaning the Converter Kit 

The 

Converter Kit

 can be cleaned with disinfectants containing 

Alcohol 50%/60%, e.g. Meliseptol 

Alkylamine 0,7% (in 2% solution), e.g. Incidin Plus 2% 

Active Oxygen 1,35% (in 3% solution), e.g. Perform 

Aldehyde 1% (in 4% solution) 

Chloride 0,4% (in 4% solution) 

 
How to clean the 

Converter Kit

 

1. 

Shut down the system by unplugging the mains power. 

2. 

For  cleaning  and  disinfection,  use  a  moist  cloth  and 
clean all surfaces using a surface disinfectant. Follow the 
disinfectant manufacturer’s recommendations. 

3. 

Carefully  clean interfaces, ensuring  that  no  liquids enter 
the system. 

4. 

For  cleaning  of  all  other  devices  refer  to  the  respective 
Instruction for Use. 

 

Storing the Converter Kit 

Use the delivered transportation bag to store or transport the 

Converter Kit

.  

 
How to store Converter Kit 
1. 

Unplug the mains power and disconnect all video cables. 

2. 

Clean the 

Converter Kit

 and put it into the transportation 

bag. 

3. 

Coil  up  each  cable  and  put  them  into  the  transportation 
bag. 

 

Trouble Shooting 

Problem 

Possible Cause 

No Power 

Check that the mains power cabling is 
correctly connected to power supply. 

No Image 

Check that all video cables are correctly 
connected. 

 
 

Expected Service Life 

Brainlab provides a minimum of eight years of service for 
devices. 

 

 

 

 

 

 