Display Reference
35
Symptoms
Possible Causes
Corrective Actions
Systolic
display
shows software
version number (e.g.,
1.53) and
Pulse
and
Reading
displays
show varying
numbers.
BpTRU is set to BP
Zero Calibration Check
or BP Reference Gauge
Calibration Check.
Exit BP Zero
Calibration Check or
BP Reference Gauge
Calibration Check by
turning power off and
then on again.
Monitor enters low-
power state often or
turns off automatically
( BPM200 & BPM300).
Insufficient charging.
Charge battery for 12
hours.
End of battery life
Replace battery (part #
BPM20036).
Battery charging
indicator remains on
for longer than 24
hours during charging.
End of battery life.
Replace battery.
BP System Errors (E0 to E15)
If a BP system error occurs during operation, an error
code will be displayed in the
Systolic
display only. All
other displays remain blank.
Press the
BP Start
button to resume measurements.
Code
Description
Corrective Action
_ _ _
User-cancelled
measurement.
None required.
E0
a
Too few pulses
detected.
Check patient for pulse. Tighten cuff
on arm and check BpTRU
for leaks in
tubing or cuff.
E07
Loss of calibration
data.
Contact BpTRU Customer Support.
Summary of Contents for BPM100
Page 2: ......
Page 14: ...Manuale di istruzioni BpTRU 8...
Page 20: ...BpTRU Operator s Manual 14...
Page 30: ...BpTRU Operator s Manual 24...
Page 48: ...BpTRU Operator s Manual 42...
Page 62: ...BpTRU Operator s Manual 56...
Page 82: ...BpTRU Operator s Manual 76...
Page 90: ...BpTRU Operator s Manual 84...