Page 1: ...EN Instruction Manual ErgoLAP BIPOLAR...
Page 2: ......
Page 3: ...to during use e g with machine cleaning likewise the specifications of the rinsing machine manufacturer must be taken into consideration Explanation of symbols WARNING information WARNING 1 Indicates...
Page 4: ...n endoscopy trocars optics electrosurgical generators etc inserted through natural or surgically made entries Intended for intermittent use under normal conditions for a period of less than 60 minutes...
Page 5: ...d medical personnel in medical establishments Before using the products carefully read the Instruction Manual 1 3 Terminology and identification of parts A Jaw B Inner shaft tube C Outer shaft tube D...
Page 6: ...lar HF surgery coagulation of tissue is effected by the application of a high frequency alternating current thereby generating heat The HF current flows from one branch of the instrument via the biolo...
Page 7: ...he product is originally packed by BOWA delivery unsterile Reading the Instruction Manual Before use read this Instruction Manual thoroughly Read the instruction manuals of other products used during...
Page 8: ...asp HF cable only by the connector when plugging in and unplugging Active electrodes WARNING Risk of unintended burns Never use worn out or faulty active electrodes Dispose of worn out or faulty activ...
Page 9: ...es Even explosions are possible through detonation Causes can include Ignition of skin cleaning solution and disinfectant Ignition of anaesthetic gases Ignition of insufflated gases Only insufflate in...
Page 10: ...tput power due to insufficient coagulation without fail check Perfect contact of all HF cables and connectors Foot switch function Insulation of the HF cable active electrodes trocar sleeve or shaft C...
Page 11: ...erences page 18 Use only sterilised products in perfect condition Before using each time carry out a thorough visual inspection and functional check Inspecting products page 12 Attach HF cable Introdu...
Page 12: ...excessive force and where possible even broken off After every procedure inspect the jaw All parts must be present Inspecting products page 12 3 4 After use Decontaminate Preparation page 8 3 5 Inspe...
Page 13: ...sible damage fractures breaks B Inner shaft tube Check straigtness C Outer shaft tube Check straigtness D Handle D1 Push button red D3 Sealing black Install D1 and D3 Check Does D1 operate smoothly Di...
Page 14: ...ch Step 4 Introduce A Push A in until audible latching occurs If necessary press D1 slightly Latching noticeable Align A as required For this turn A until further latching is heard To facilitate this...
Page 15: ...he instrument immediately after use or no later than 2 hours after use Use only aldehyde free disinfectants recommended Gigasept Instru AF designed for the disinfection of instruments e g DGHM or FDA...
Page 16: ...e for the designated products Effective cleaning proven by BOWA Machine cleaning 90 C 5 minutes using an alkaline cleaning agent with surfactant additive Neodisher MediClean forte The manufacturer acc...
Page 17: ...toring the device Ambient conditions Operation and storage Transport Temperature 10 C to 40 C 20 C to 50 C Relative humidity 0 to 75 non condensing Atmospheric pressure 500 to 1060 hPa Selection of re...
Page 18: ...033 750 010 Handle 750 000 Outer shaft tube 755 033 Inner shaft tube 756 033 Jaw 751 033 752 033 753 033 754 033 760 033 761 033 Connecting Cables 101 245 4 5 m 287 245 4 5 m 351 245 4 5 m 350 245 4...
Page 19: ......
Page 20: ...gn changes Copyright by BOWA electronic Gomaringen Germany BOWA electronic GmbH Co KG Heinrich Hertz Strasse 4 10 D 72810 Gomaringen Germany Phone 49 0 7072 6002 0 Fax 49 0 7072 6002 33 info bowa medi...