8 Reprocessing
38
BOWA-IFU-MN031-554-ERGO315R-S3-EN-20210219
8.5.2.
Placing the parts in the reprocessing basket
NOTE
Risk of damage to the instrument components from
incorrect placement in the CDM!
Use the reprocessing basket for cleaning to avoid
damage to the instrument components.
Ensure that the HF cable is not kinked or pinched.
Risk of damage to the instrument components when
putting them in the reprocessing basket or removing them
from the reprocessing basket
Avoid bending the instrument components when
clipping them into the holders. Press the instrument
components into the holders with equal pressure on
both ends.
1.
Connect the cleaning hoses of the CDM to the Luer Lock
couplings of the cleaning adapters.
2.
Place the instrument components with the attached cleaning
adapters in the corresponding recesses in the reprocessing
basket as indicated by the symbols.
3.
Feed the hoses out of the reprocessing basket through the
openings in the end of the reprocessing basket provided for that
purpose.
Ensure that:
•
The instrument components are positioned to ensure
exposure to the cleaning media