6 Operation
BOWA-IFU-MN031-554-ERGO315R-S3-EN-20210219
25
6.2.
Function test in the operating theatre
1.
Check whether the electrodes can be easily opened and closed
with the handle.
2.
Lock the handle by pressing the actuation lever.
3.
Actuate the blade trigger and check it for ease of movement.
4.
Check the activation by pressing the handswitch.
The activation signal sounds during activation of the
handswitch.
6.3.
During surgery
Risk of patient injury due to tissue clamping, especially
with limited visibility! Prepare the tissue to be sealed so it
is as free as possible in order to avoid unintentional
clamping.
Only carry out the operation with adequate visibility.
Risk of patient injury due to hot electrode surfaces and
steam emission!
Instrument tips may still be hot immediately after HF
power has been switched off.
Maintain sufficient distance between instrument tips and
sensitive tissue structures, such as nerves, pancreas or
intestines.
Ensure that hot vessel sealing instruments are not used
for preparation.
Do not lay the vessel sealing instrument on the patient.
Introducing the instrument
1.
Using the handle, close the electrodes.
2.
Insert the instrument into the trocar sleeve.