2 Safety
900-400_IFU_V1.0.0.6.17.9_20140-S2-20130604-EN
Operating Manual ARC 400
11
2.3.
Personal safety instructions
2.3.1. Ambient conditions
Do not use the HF device in the immediate vicinity of the patient. Observe the
minimum distances recommended by BOWA, as shown in the following figure.
2.3.2. Patients with pacemakers
Malfunction or destruction of the pacemaker can endanger the life of the patient or
result in irreversible injuries to the patient.
In the case of patients with pacemakers, consult the cardiologist before carrying
out HF surgery.
Use bipolar HF methods.
Attach the HF neutral electrode close to the operating field.
Set the demand pacemaker to a fixed frequency.
Ensure that the pacemaker does not come into contact with the HF electrode.
Keep a fully operational defibrillator within reach.
Carry out a postoperative pacemaker check.