De‑linking a Remote Control from a Stimulator
To de‑link a Remote Control from a Stimulator:
1
Press and hold the
P
button to unlock the Remote Control.
2
Enter the Clinician’s screen on the Remote Control by pressing and holding both
the and
P
buttons simultaneously.
3
Scroll down to “Clear Link” and select “Clear Link” by pressing the
P
button.
4
Confirm that you want to clear the link by pressing the
P
button or cancel by
pressing the button.
5
Enter the password A‑B‑C.
Note:
To re‑link the Remote Control, see the “Linking a Remote Control to a