45
Parameter
Value
Temperature
50 ºC (122 ºF) minimum and 60 ºC (140
ºF) maximum
Humidity
40% minimum and 80% maximum; non-
condensing
EO dwell time
2 hours
EO concentration minimum
450 mg/L
Minimum aeration time
12 hours at 60 ºC (140 ºF)
Number of sterilization cycles permitted
6395 Telemetry Wand = 25
6763 PSA cable = 50
Steam Sterilization Instructions
•
Prior to sterilization, thoroughly clean the Model 6763 PSA cable as
indicated in the "Cleaning Cables and Wands" on page 44.
•
Follow the recommendations of the steam sterilization equipment
manufacturer and allow the specified drying time to fully elapse prior to
use.
•
Only FDA-cleared pouches should be used during sterilization processing.
Only the Model 6763 PSA cable may be steam sterilized.
Parameter
Value (gravity sterilizers)
Value (dynamic-air-
removal sterilizers)
Temperature
121 ºC (249.8 ºF)
132 ºC (269.6 ºF)
Steam dwell time
30 minutes
4 minutes
Drying time
9
15 minutes
20 minutes
Number of sterilization
cycles permitted
50
50
Battery Status, Installation, Replacement, and
Recycling
The Programmer battery has been tested and approved for hospital and clinic
use. Battery status is a percent of charge remaining (see Figure 28 Battery
Status Icons Indicating Charge Percentage on page 46) and is displayed in the
9.
This represents the typical drying time. Once sterilized, ensure that the device is dry prior to
removing it from the chamber.