Electrosurgical Cables
Electrosurgical cables must be kept at least 30 cm (12 in) away from the
LATITUDE Programming System to avoid false screen traces when
electrosurgical energy is applied.
Figure 7.
Electrosurgical Cables Distance from LATITUDE Programming System
Prepare for ZIP (RF) Telemetry
NOTE:
The ZIP telemetry feature is not available for all Boston Scientific PGs.
For more information, refer to the associated product literature for the PG being
interrogated.
For PGs that communicate using ZIP telemetry:
1.
For optimum ZIP telemetry communication, position the LATITUDE
Programming System within 3 m (10 ft) of the PG.
2.
Remove obstructions between the LATITUDE Programming System and
the PG.
NOTE:
Reorienting or repositioning the LATITUDE Programming System
may improve ZIP telemetry performance.
NOTE:
The Model 3203 S-ICD Telemetry Wand may be used as a third RF
antenna to improve RF telemetry performance.
NOTE:
If ZIP telemetry performance is not successful, use the Model 6395
Telemetry Wand to interrogate the PG.