![Boston Scientific Latitude 3300 Operator'S Manual Download Page 27](http://html.mh-extra.com/html/boston-scientific/latitude-3300/latitude-3300_operators-manual_2796085027.webp)
21
performance. Refer to "Steps to Improve ZIP (RF) Telemetry Performance"
on page 23 for additional information.
Make Physician Side Connections
As needed, make the following connections on the left side of the LATITUDE
Programming System.
[1] Power (on/off) button (light gray) [2-4] USB 2.0 ports (dark gray) [5] USB 3.0 port (blue) [6]
Ethernet port (orange) [7] DisplayPort out (red-orange) [8] DC power connection for Model 6689
Power Adapter (green)
Figure 6.
Left Side (Physician) Panel
1.
Connect the power cord to the DC receptacle on the left side panel of the
Programmer.
2.
To connect an external USB printer, attach the appropriate USB cable (2.0
or 3.0) to the appropriate USB port on the Programmer. Then, ensure that
the printer is connected to external power.
NOTE:
Connect the printer to the USB port, then wait 30 seconds for the
system to recognize the printer before sending files to the printer.
NOTE:
The LATITUDE Programming System has Bluetooth
®
capability,
which can be used to connect with Bluetooth
®
capable printers. Refer to
the Network and Connectivity Operator’s Manual (Model 3924) for
additional setup and use Information.
3.
Use the DisplayPort Out connector to attach an external monitor. Then,
ensure that the monitor is connected to external power.
4.
To connect to a LAN, attach an Ethernet cable to the Ethernet port.
NOTE:
Connect the Ethernet cable only to the RJ45 Ethernet port
connector on the Model 3300 Programmer.
NOTE:
Additional steps need to be completed when using Bluetooth
®
or
LAN communications. Refer to the Network and Connectivity Operator’s
Manual (Model 3924) for additional information.
5.
Ensure the power adapter cable is plugged into the DC port on the left side
of the Programmer and the power cord is plugged into the power adapter.
NOTE:
Ensure the left side of the device is accessible at all times so that