User guide

Rayplicker Cobra

8. 

ACHIEVING A GOOD SHADE TAKING

8.1. 

Good practices for using the Rayplicker Cobra

The specific design of the Rayplicker Cobra measuring head is the result of an ergonomic study 

which gives it accessibility right up to the pre-molars as well as trouble-free handling thanks to its 

wireless technology. For optimal shade-taking results, a few good practices listed below should 

be followed.

MOUTH CLOSED

TOOTH CENTERED 

INCORRECTLY

TIP INCORRECTLY 

POSITIONED

EXCESSIVE PRESSURE 

ON THE GENGIVA

CONDENSATION

WRONG ANGLE

VISIBLE LIP

WRONG INCLINATION

BEST PRACTICES FOR USING RAYPLICKER COBRA

INCORRECT PRACTICES

GREY EDGES PARALLEL TO SCREEN EDGES

MOUTH OPENED AND BREATHING THROUGH THE NOSE

SALIVA LIMITATION

CENTERING THE TOOTH, WITHOUT ANGLE NOR INCLINATION

TIP UNDER THE LIP

TIP IN CONTACT WITH ADJACENT TEETH

TIP IN CONTACT WITH THE GINGIVA

DON’T MOVE DURING THE ACQUISITION (1s)

CALIBRATION TIPS

DAMAGED TIP

Internal aspect and/or

external shiny and sticky

FAILURE TO RESPECT THE DISINFECTION AND 

STERILIZATION PROCEDURES CAUSES A HASTY 

DEGRADATION OF THE TIPS.

A DAMAGED CALIBRATION TIP (DISCOLORATION, 

MODIFIED ASPECT,...) MUST BE CHANGED. 

LIFETIME VERIFIED FOR 100 AUTOCLAVE CYCLE PASSES. 

NEW TIP

Internal aspect and 

external gray and matte

Summary of Contents for Rayplicker Cobra

Page 1: ...Rayplicker Cobra USER GUIDE Version 1 0 11 2022 Ref UG RPC EN 1 0...

Page 2: ...te 11 2022 Compatible with Rayplicker Cobra spectrophotometer Copyright BOREA SAS This document is the property of BOREA SAS and may not be used reproduced or communicated to third parties without its...

Page 3: ...Turning on the device 13 5 DEVICE INTERFACES 13 5 1 User interface navigation 13 5 2 Note on device illustrations 14 5 3 Rayplicker Cobra interfaces tree 14 5 4 Wifi menu 15 5 5 Information menu 16 6...

Page 4: ...ection and sterilization procedures 23 10 TECHNICAL SPECIFICATIONS 24 10 1 Identification 24 10 2 AC adapter 24 10 3 Handpiece 24 10 4 Battery 24 10 5 Environmental characteristics 25 10 6 Regulation...

Page 5: ...t shade guides used to determine color as well as on the basis of CIE L a b values Precise and ergonomic the Rayplicker Cobra solution is also equipped with a software allowing the processing and reco...

Page 6: ...pproved by BOREA Protect the device and charger against damage Environmental conditions Do not store your device in hot or cold places Do not expose the device to temperatures below 0 C 32 F or above...

Page 7: ...is not in use do not leave it plugged into a power outlet Only use battery packs chargers accessories and supplies approved by BOREA Please only use the rechargeable battery packs supplied with the d...

Page 8: ...The Rayplicker Cobra device is a Class I medical device It bears the CE marking in accordance with European Regulation EU 2017 745 relating to medical devices Date of first affixing of CE marking 202...

Page 9: ...incorporate a chromatic reference gray pattern necessary for the proper functioning of the device through a self calibration step not perceived by the user An oled screen is integrated into the devic...

Page 10: ...battery pack 1 EA0207 Docking station 1 SF0210 Dongle 1 RA0201 Main charger 5V 2 4A 1 EA0211 Documentation Quick Start Guide Disinfection and Sterilization Procedures Best practices for using Rayplic...

Page 11: ...ssories are available from BOREA SAS or your distributor ACCESSORIES 6 Calibration tips Autoclavable Ref RA0209 Main charger 5V 2 4A Ref CA0204 Wireless battery pack Ref EA0207 Docking station Ref SF0...

Page 12: ...clean and in good condition It must also be protected from the sun s rays and any other type of external aggression The condition of the tip attests to the quality of the calibration and guarantees t...

Page 13: ...s tree page 14 The strength of the Wifi signal is indicated at the top right of this screen 5 DEVICE INTERFACES The Rayplicker Cobra device is equipped with an OLED touch screen To navigate between th...

Page 14: ...Rayplicker Cobra interfaces tree Scroll left Scroll right Back Device connected but not synchronized Device connected but not synchronized Don t move Transfer in progress Acquisition validated Error...

Page 15: ...onnected to the Rayplicker Vision software This interface only shows the device name This can be customized from the Devices settings in the Rayplicker Vision software This interface indicates the pas...

Page 16: ...r guide Rayplicker Cobra Here is displayed the battery level of your device 20 20 40 5 5 Information menu 60 80 100 Here is displayed your device serial number Here is displayed your device software v...

Page 17: ...s Step 1 By default after thirty seconds of inac tivity the Cobra locks If this symbol is displayed press the trigger to access the device functions again 2nd step The system automatically shuts down...

Page 18: ...reate the patient file of your choice on the Rayplicker vision software Make sure your device is connected and synced A video stream is launched to help you po sition the appliance correctly in contac...

Page 19: ...g the sterilizable tip which has a gray pattern It is necessary to ensure the cleanliness and condition of this calibration tip It must also be protected from the sun s rays and any other type of exte...

Page 20: ...error occurred during acquisition Please redo the shade taking An error occurred during the transfer Click on CANCEL to resume a shade Another Rayplicker Vision software within your structure may be...

Page 21: ...WRONG ANGLE VISIBLE LIP WRONG INCLINATION BEST PRACTICES FOR USING RAYPLICKER COBRA INCORRECT PRACTICES GREY EDGES PARALLEL TO SCREEN EDGES MOUTH OPENED AND BREATHING THROUGH THE NOSE SALIVA LIMITATI...

Page 22: ...ng type of drying drying time for soaking bath type products dilution soaking time type of rinsing type of drying drying time renewal For sterilization autoclave at 134 C for 20 min Prion cycle Lifeti...

Page 23: ...d al virucidal and tuberculocidal properties Please follow the manufacturer s instructions contact time required See below for risk areas This device disinfection protocol is mandatory after each use...

Page 24: ...ncy 50 60 Hz Output voltage 5 0V DC Output current 2 4 A Weight g 155 Dimensions mm L 185 x l 40 x H 30 Sensor CMOS 648 x 648 px Spectrophotometer Measuring range 400 700 nm Measuring head dimensions...

Page 25: ...ation of its declarations of placing on the market with ecosystem 10 6 Regulation and standardization The use of the Rayplicker Cobra device at the same time or in close proximity to another electroni...

Page 26: ...y Characteristic of a material that allows light rays to pass and undergoes scattering You can see through translucent material but you cannot distinguish shapes 11 3 Patents 11 4 Warranty The Rayplic...

Reviews: