10.1 List of error, malfunction and generated messages
Problem
desription
Likely cause
Action to be taken by
the operator to solve
the problem
Actions to be taken
by the qualified
service personnel
The bed is not
responding to
controller
commands.
The bed probably does
not have electric
supply and the battery
is already depleted.
Plug the power supply
cord to the electric socket
(full recharge of the
battery takes 8 hours).
Check, if the electric
network is on and the
socket in function. Use
other electric socket.
The hand
controller or
some of its
buttons are not
working.
The hand controller
cable is not properly
inserted in the bed
connector.
Make sure, the hand
controller cable is
inserted properly in the
bed connector.
Check, if the bed
responds with another
hand controller. If so,
replace the hand
controller.
The function lock
indicator is
flashing and after
pressing any
button on the
bed, a short
alarm signal
sounds
Emergency stop
device is not working
Calling service
Replacing the DPS
controlboard or erasing
the error flag using ‘r’
sent via diagnostics
When all
functions are
actively locked,
after pressing
any button, a
short acoustic
signal sounds
Locked buttons
Unlocking buttons of
keyboards by holding
down the lock button
-
The bed emits
intermittent
acoustic signal
and one of the
controllers is
not functioning
Controller failure
Call the service
Repair of the faulty
controller