MS-301 Multi parameter Water Quality Sensor
Web:www.boquinstrument.com
calibration according to actual requirement to make integrated multi parameter water quality sensor
immerge into standard solution, and at the same time, input corresponding standard solution value in the
square frame of the hand manipulator. Then, choose the key of “calibration”, and press the key of “OK”.
At last, press the key of “Enter”.
6.2.12 Ammonia Nitrogen Calibration
Get into the interface ammonia nitrogen calibration, and press the key of “calibration”. Press the key of
“Enter” to calibrate the ammonia nitrogen. Choose first point/second point/third point/fourth point/fifth
point ammonia nitrogen calibration according to actual requirement to make integrated multi parameter
water quality sensor immerge into standard solution, and at the same time, input corresponding standard
solution value in the square frame of the hand manipulator. Then, choose the key of “calibration”, and
press the key of “OK”. At last, press the key of “Enter”.
Chapter 7 Maintenance
Danger