Operating Manual of Ammonia Nitrogen Sensor
31
Chapter 7 Communication Protocol
A. The sensor probe has Modbus RS485 function, communicates in RTU mode, and the default baud rate
is 9600. It can communicate with the host computer (PC/digital instrument, etc.) with RS485
communication function. The specific MODBUS-RTU table is shown in the table below.
MODBUS-RTU
Baud rate
9600/4800/19200/38400/115200
Data bit
8 digits
Parity
no
Stop bit
1 digit
Slave address
1-254 (default 10)
B
、
Adopt MODBUS standard communication protocol. The specific agreement content is shown in the
table below.
Address Name
Data
Type
Operate
Type
Remarks
0
Temp. Value
float
read
03 command
2
pH Value
float
read
03 command
16
Ammonia Nitrogen
float
read
03 command
18
Potassium ion value
float
read
03 command
35
Equipment serial number SN1
unsigned read
03 command
36
Equipment serial number SN2
unsigned read
03 command
37
Equipment serial number SN3
unsigned read
03 command
70
Year
float
read/write
03 command /16 command
72
Month
float
read/write
03 command /16 command
74
Day
float
read/write
03 command /16 command
76
Hour
float
read/write
03 command /16 command
78
Minute
float
read/write
03 command /16 command
80
Second
float
read/write
03 command /16 command