Otherwise, vaccine will be spilled over the device while a bottle is being connected to or disconnected 

from the bottle mount.  

If using the tubing collar to connect a larger bottle via tubing, 

prefill the tubing

 (!) prior to attaching it 

to the tubing collar. When the tube starts to fill, kink the tube just before the end to prevent loss of 

vaccine. Then quickly connect the tubing to the adapter. 

 

6.

 

Device Start 

Once the vaccine bottle is connected, insert the battery and the device will turn on, showing both the 

Boehringer Ingelheim and the FreVAX™ logos on the display. Both logos are shown for 1 second. 

     

 

If the battery is sufficiently charged:

 The display will switch initially to the User Menu, unless the 

device is in a loaded condition (cylinder filled with vaccine / rinsing fluid) – in which case the display 

will show “! Loaded Device ! Press any button” and the LED bars turn red.  

 

By pressing one of the display buttons, the device will rinse once, to 

unload the device. The liquid will be sprayed out through the injection 

head. 

 

Afterwards, the device will switch to the initial display of the User Menu (see chapter 7).  

Before pushing the trigger for the first time, make your settings in the User Menu. 

If the battery is too low

: The display will show “Low Battery Please 

Charge”.  

 

 

Summary of Contents for FreVAX 101

Page 1: ...operating and cleaning 4 4 General description and information 5 4 1 Description 5 4 2 Packaging contents 5 4 3 Operating requirements 5 4 4 Target species 5 4 5 Mode of action 5 5 Bottle connection...

Page 2: ...Rinsing 13 10 2 Cleaning of the device 14 11 Maintenance and Repair 14 11 1 Cylinder Service 14 11 2 Maintenance and repair service 16 12 Dismantling 18 13 The battery and the charger 19 14 Troublesho...

Page 3: ...lly and follow the instructions copyright Henke Sass Wolf GmbH 8300053415 10 20 Subject to amendments in technology and design Printed in Germany The FreVAX injector is manufactured by Henke Sass Wolf...

Page 4: ...other people In case of accidental injection of vaccine and or rinsing fluid seek immediate medical advice and show the vaccine information sheet Beware of the spike in the bottle holder It may contai...

Page 5: ...bottle for use with defined cleaning liquids 5 Universal bottle collars 20mm 30mm 6 Tubing adapter 7 Tubing under the device Threaded bottle collar for use with PET bottle 8 Lubricant for cylinder ser...

Page 6: ...Operation of the FreVAX begins with inserting a bottle of vaccine into the bottle holder A and pressing the trigger F to activate the injector This will start the pump and vaccine is automatically dra...

Page 7: ...ector on a solid surface as shown on the picture and push the bottle onto the spike using the pre punctured hole CAUTION Without pre puncturing of the bottle stopper rubber particles may be drawn into...

Page 8: ...and the FreVAX logos on the display Both logos are shown for 1 second If the battery is sufficiently charged The display will switch initially to the User Menu unless the device is in a loaded conditi...

Page 9: ...re the first injection The fifth page shows the Service Counter which is counting from 12 000 to 0 If the value falls below 0 the counter continues counting negatively Press and hold the left display...

Page 10: ...vaccinating them with min time lapse between applying the device to the skin of the piglet and the injection itself Press and hold the left display button to select the alternative setting Delay s whi...

Page 11: ...Afterwards only rinsing is possible indicated by blue flashing LED bars until the air exits at the injection head This takes a minimum of 3 rinsing cycles as the liquid circuit has a volume of a littl...

Page 12: ...press and hold the trigger to load the device for the first injection During loading Load Device is displayed After loading the device is ready to vaccinate which is indicated by a white flashing LED...

Page 13: ...ce from the skin of the piglet to enable the next loading cycle and injection When kept in contact with the piglet s skin after an injection and still reloading the device will show Please release Noz...

Page 14: ...mercially available disinfectant while wearing rubber gloves and turning moving the safety sleeve around the injection head to remove all traces of dirt Clean the grey housing and the battery either b...

Page 15: ...the head facing downwards in order to drain residues of vaccine or rinsing fluid Step 2 Remove old cylinder Unscrew the old cylinder with a 15 mm wrench Hold the device with the head facing downwards...

Page 16: ...ul completion of cylinder service please reset service counter Starting from the initial screen push the right button three times until reaching menu point Service Counter Press button Reset for at le...

Page 17: ...le during loading to increase suction pressure Farm User 3 If both doesn t help repair through manufacturer Henke Sass Wolf Position sensor 1 Optionally use device without position sensor Farm User 2...

Page 18: ...The FreVAX device may only be dismantled by authorized service engineers In case of unauthorized dismantling the device will lose its warranty Do not open the FreVAX device by yourself due to health...

Page 19: ...ecommended that batteries are not recharged before falling below 50 loading capacity Never store empty batteries as this could damage them and make them unusable Never send ship defective batteries to...

Page 20: ...ions LED lights orange Battery is empty and charging LED lights yellow Battery is partly charged and still charging LED lights green Battery is fully charged The battery charger is not waterproof Oper...

Page 21: ...ot of vaccine on the skin of the animal and no puncture visible as mentioned in chapter 9 it is normal that a small residue of vaccine will remain on the piglet s skin Air inside the tubing system Rin...