User Guide e-CELSIUS Performance.2
EN
21
6.7.5
Detailed temperature data visualization
To access to a detailed visualization, select the capsule by positioning it in the right of the
screen. Then press the OK button, a new screen appears with the following information:
-
The number of the selected capsule (from 1 to 3).
-
The last temperature data collected (°C).
-
The delay since the last temperature data collected (h:mm:ss).
-
The minimal and maximal value and the battery status of the selected capsule.
-
Synchronisation indicator C1, C2, C3 link to relative pills.
In this display mode, the screen never turns automatically in standby mode, thereby
facilitating continuous data visualization of the capsule. This display mode slowed the
synchronization of data between the monitors and associated capsules. It is
recommended to leave this display mode to optimize the speed of data recovery.
6.7.6
Synchronization of the data in memory of the capsule
It is recommended to be vigilant in environments with high metal stress (reinforced concrete
wall ...) and to regularly check on the monitor screen that communication with the capsule is
not disturbed.
In the screen "Data View ", a star (*) associated with a capsule number indicates that the
monitor has to be synchronized with the capsule. An orange star means that the last data have
been collected more than 5 minutes.
The capsule has an internal memory that automatically records the last 2000 measurements
collected. When the communication between the monitor and / capsule(s) is disrupted
(indicated by the star on the temperature display screen), the monitor is not receiving / the
data.
Nevertheless, there is a feature in the monitor that automatically recovers the missing data as
soon as communication is restored.
The monitor synchronizes also automatically its data with the data available in the
capsule memory. The number of data for which the synchronization will be performed
has to be previously set during the configuration of the monitor. The number of data to
synchronize between the monitor and the capsules can be set between 0 and 2000. This
Picture 20: Detailed visualization screen of a capsule