Operator’s Manual Simple AFG Application
Berkeley Nucleonics Corporation
[email protected] | www.berkeleynucleonics.com
2955 Kerner Blvd, San Rafael, CA 94901 | 800-234-7858
Rev 1.1
Trigger
In Settings
The Trigger In settings parameters are located in the Device Setting page. These parameters are shared
by all channels of the instrument.
Trigger In Settings: 2 Channels Model
Trigger In Setting
Description
Source
Button:
the Trigger event is provided to the instrument by pressing the
Trigger button on the keyboard or the Trigger button on the menu toolbar
or by issuing a trigger by Remote Command.
Timer:
the Trigger is internally generated by a Timer. The Timer count
interval is set by Interval [s] textbox.
External:
a Trigger event is generated by the signal applied externally to
the Trigger In 1 SMA connector when it crosses the selected Threshold
with the selected Slope. The Threshold value and the Slope can be defined
in the relative textbox and slider.
Note:
the Trigger buttons and the Trigger from remote command are
always active, independently from the selected Trigger Source.